- Check ballast module power and the wiring harness
- Inspect the lamp and the lamp connector pins
- Replace the UV lamp
- If unchanged — replace the ballast module
- Clean the quartz sleeve and the sensor window
- Check the UVIS cable and connector
- Replace the UV sensor (OPB000100)
- Calibrate against the master sensor kit
- Check cabinet ventilation and filters
- Clean the cooling fins
- Verify ambient temperature is within spec
- Replace the ballast module
- Cut power and isolate the system
- Read and export the controller error log
- Contact Atlantis-UV — do not restart before clearance
- Check lamp age against the 6,000 h interval
- Clean the quartz sleeve
- Measure UVT — check upstream water quality and turbidity
- Isolate and depressurise the reactor
- Inspect the sleeve O-ring (MP0023000) and sleeve holder O-ring (MPE007500)
- Re-seat or replace the seal, then run the return-to-service checklist
- Verify water flow is actually present
- Check the flow sensor wiring and the 4–20 mA loop
- Confirm the interlock configuration on the controller
- Check the RS485 wiring and termination
- Verify the Modbus address and baud rate
- Restart the controller
Service checklists
Open a checklist, tick the items as you go, fill in the site details at the bottom and send the completed report to Atlantis-UV — or print it as a PDF for your own logbook. Your progress is saved in this browser as you fill it in, so you can close the page and come back. It is cleared once the report is sent.
Recommended site spare set
What every operating plant should hold on site so a lamp or seal failure never becomes downtime. One set per installed reactor — quoted as a package on request.
Requests
Everything reaches the Atlantis-UV service team directly. We answer within one working day; for a system that is down, call +421 918 101 507.
Manufacturer sources
Atlantium Technologies is the manufacturer of the HOD UV systems we supply and service. Datasheets, approvals and application material come from these pages.